The calibration
services provided by the metrology laboratories of QSAE are in the
following measurement fields.
Dimensional
measurement
Mass measurement
Volume measurement
Density measurement
Pressure measurement
Force measurement
Temperature measurement
Electrical measurement
Frequency measurement
Dimensional
measurement

|
Item No
|
Measured quantity or
type of measuring instrument
|
Range
|
|
1
|
Gauge blocks
|
0.5 to 100 mm
|
|
2
|
Micrometers
|
0 to 100 mm
|
|
3
|
Vernier calipers
|
0 to 300 mm
|
|
4
|
Dial gauges
|
0 to 40 mm
|
|
5
|
Rules
|
0 to 1000 mm
|
|
6
|
Glass polygon
|
0 to 360 degree
|
|
7
|
Special dimensional measurement
services
|
To be agreed
|
[Back to top]
Permanent Laboratory
Mass
measurement

| Measured quantity / Calibration item |
Range |
Measurement conditions / procedure |
Best measurement capability 1) |
Remarks |
| Conventional Mass |
1 mg, 2 mg, 5 mg |
Only weights with a certificate confirming the conformity with the requirements of
OIML R 111:2004 can be calibrated.
|
0,006 mg |
OIML Recommendation R111,
class F1 |
| 10 mg |
0,008 mg |
| 20 mg |
0,010 mg |
| 50 mg |
0,012 mg |
| 100 mg |
0,016 mg |
| 200 mg |
0,020 mg |
| 500 mg |
0,025 mg |
| 1 g |
0,03 mg |
| 2 g |
0,04 mg |
| 5 g |
0,05 mg |
| 10 g |
0,06 mg |
| 20 g |
0,08 mg |
| 50 g |
0,10 mg |
| 100 g |
0,16 mg |
| 200 g |
0,3 mg |
| 500 g |
0,8 mg |
| 1 kg |
1,6 mg |
| 2 kg |
3,0 mg |
| 5 kg |
8,0 mg |
| 10 kg |
16 mg |
[Back to top]
Volume
measurement

| Item
No. |
Measured quantity
or type of measuring instrument |
Range |
| 1 |
Pipettes and burettes |
All |
| 2 |
Measuring cylinders and flasks |
All |
| 3 |
Density bottles |
All |
| 4 |
Flow/bulk meters |
All |
| 5 |
Storage tanks |
Up to 100 000 L |
| 6 |
Road tankers |
All |
| 7 |
Special volume measurement
services |
To be agreed |
[Back to top]
Density
measurement

| Item
No. |
Measured quantity
or type of measuring instrument |
Range |
| 1 |
Hydrometers |
600 kg/m3 to 2000 kg/m3 |
| 2 |
Special density measurement services
|
To be agreed |
[Back to top]
Pressure measurement

| Item
No. |
Measured quantity
or type of measuring instrument |
Range |
| 1 |
Pressure balances |
Up to 60 MPa |
| 2 |
Pressure gauges |
Up to 60 MPa |
| 3 |
Barometers |
800 mbar to 1070
mbar |
| 4 |
Special pressure
measurement services |
To be agreed |
[Back to top]
Force
measurement

| Item
No. |
Measured quantity
or type of measuring instrument |
Range |
| 1 |
Tensile testing machines |
Up to 500 kN |
| 2 |
Compressive testing machines |
Up to 500 ton |
| 3 |
Load rings |
Up to 500 kN |
| 4 |
Special force measurements
services |
To be agreed |
[Back to top]
Permanent Laboratory
Temperature
measurement

| Temperature
Resistance thermometers
|
-40 °C
|
to |
30 °C |
Comparison with a standard resistance thermometer |
20 mK |
Alcohol bath |
| > 30 °C |
to |
80 °C |
30 mK |
Water bath |
| > 80 °C |
to |
200 °C |
100 mK |
Oil bath |
| > 200 °C |
to |
660 °C |
600 mK |
Fluidized bath |
| Thermocouples,
noble metal
|
0 °C |
to |
200 °C |
Comparison with resistance thermometer |
0,5 K |
Liquid bath |
| > 200 °C |
to |
600 °C |
0,7 K |
Fluidized bath |
| > 600 °C |
to |
900 °C |
Comparison with thermocouple |
1,1 K |
|
| > 900 °C |
to |
1200 °C |
1,5 K |
Furnace |
| Thermocouples,
base metal
|
0 °C |
to |
200 °C |
Comparison with thermocouple |
1,0 K |
Furnace |
| > 200 °C |
to |
600 °C |
1,0 K |
| > 600 °C |
to |
900 °C |
1,5 K |
| > 900 °C |
to |
1200 °C |
2,0 K |
| Liquid in glass thermometers |
-40 °C |
to |
30 °C |
Comparison with a standard resistance thermometer |
25 mK |
Alcohol bath |
| > 30 °C |
to |
80 °C |
30 mK |
Water bath |
| > 80 °C |
to |
200 °C |
100 mK |
Oil bath |
| >200 °C |
to |
360 °C |
600 mK |
Fluidized bath |
| Direct reading electrical thermometers |
-40 °C |
to |
30 °C |
Comparison with a standard resistance thermometer |
20 mK |
Alcohol bath |
| > 30 °C |
to |
80 °C |
30 mK |
Water bath |
| > 80 °C |
to |
200 °C |
100 mK |
Oil bath |
| > 200 °C |
to |
660 °C |
600 mK |
Fluidized bath |
| Mechanical (dial) thermometers |
-40 °C |
to |
80 °C |
Comparison with a standard resistance thermometer |
30 mK |
Alcohol or Water Bath |
| > 80 °C |
to |
200 °C |
100 mK |
Oil Bath |
| >80 °C |
to |
400 °C |
600 mK |
Fluidized bath |
[Back to top]
Direct reading electrical thermometers
Electrical
measurement

| Item
No. |
Measured quantity
or type of measuring instrument |
Range |
| 1 |
DC Voltage |
Up to 1100 V |
| 2 |
DC Current |
Up to 20 A |
| 3 |
AC Voltage |
Up to 1000 V (10 Hz to 100 kHz) |
| 4 |
AC Current |
Up to 20 A (0 kHz - 5 kHz) |
| 5 |
Resistance |
0.001 to
1 M |
| 6 |
Electrical energy |
 |
[Back to top]
Frequency measurement

|
Item No.
|
Measured quantity
or type of measuring instrument
|
Range
|
|
1
|
Rubidium frequency standard
|
5 MHz
|
|
2
|
Crystal oscillator
|
5 MHz
|
|
3
|
HF Synthesizer
|
0.1 MHz - 1019 MHz
|
|
4
|
LF Synthesizer
|
0 MHz - 2 MHz
|
|
5
|
Universal counter/timer
|
0 MHz - 250 MHz
|
[Back to top]
|
|